Basic Info.
Model NO.
ICP-TY-9000
Portable
Non Portable
Certification
CE, ISO9001
Type
Full Spectrum Direct Reading Spectrometer
Wavelength Range
130 ~ 800nm
Operating Temperature
(10 ~ 35)°c
Storage Temperature
(0 ~ 45)°c
Transport Package
Stand Export Package
Trademark
Labkits
Origin
Hunan China
Production Capacity
500PCS/Month
Product Description
Item: Full Spectrum Direct Reading Spectrograph Spectrometer for Metal Analysis For Sale
Model: ICP-TY-9000
Model: ICP-TY-9000
Specification and technical parameter
Optical system
l full spectrum optical system of Palmer type - runge structure
l Maximum wavelength range: 130 ~ 800nm
l Multiple high performance CCD detector
l Integrated optical chamber process molding, resistant to change of environment temperature
l Vacuum light chamber design guarantees C, S, P, could achieve the best performance.
l light room temperature control, constant temperature is 35 °C
l Directed type optical technology and lens MgF2 materials ensure that the best energy of UV wavelength of C, S, P.
| Item | Indicator |
| Detection substrate | Fe |
| Detection time | According to the type of test sample, usually it is about 25s. |
| Optical system | Palmer type - runge Rowland circle full spectrum vacuum type optical system |
| Wavelength range | 130 ~ 800nm |
| Focal length | 400mm |
| Detector | High-performance CCD array |
| Electrode | Tungsten material spray electrode |
| Analytical gap | Sample Stage Analytical gap:3.4mm |
| Operating temperature | (10 ~ 35)°C |
| Storage temperature | (0 ~ 45)°C |
| Operating humidity | 20% ~ 80% |
| Argon purity requirements | 99.999% |
| Argon gas inlet pressure | 0.5MPa |
| Argon flow | Stimulating flow is about 3.5L/min, maintaining flow is about 0.4L/min. |
| Stimulating max power | 400VA |
| Light source type | brand new adjustable digital light source, high energy preliminary combustion science (HEPS) |
| Discharge frequency | 100-1000 Hz |
| Discharge current | Max 400A |
| Spark Stand aperture | 13mm |
| Vacuum system | Vacuum software automatically control and monitor |
| Working power | 220V AC, 50/60Hz, the single-phase power supply of protective grounding |
| Instrument dimensions | 860*720*500 |
| Instrument weight | 80kg |
Optical system
l full spectrum optical system of Palmer type - runge structure
l Maximum wavelength range: 130 ~ 800nm
l Multiple high performance CCD detector
l Integrated optical chamber process molding, resistant to change of environment temperature
l Vacuum light chamber design guarantees C, S, P, could achieve the best performance.
l light room temperature control, constant temperature is 35 °C
l Directed type optical technology and lens MgF2 materials ensure that the best energy of UV wavelength of C, S, P.